Comparison of bare metal stent and paclitaxel-eluting stent using a novel rat aorta stent model.
|
|
WRITER :
stent
HIT : 1,234
|
B37._J_Vet_Sci._2011;12_2_143-149..pdf (3.6M) [0] DATE : 2016-03-13 12:17:46 |
B37. Kwon JS, Park RK, Shim TJ, Jeong MH, Cho MC, Ahn Y, Kim DW; Comparison of bare metal stent and paclitaxel-eluting stent using a novel rat aorta stent model. J Vet Sci. 2011;12(2)143-149.
(Abstract)
The purpose of our study was to create a novel rat aorta stent implantation model. Stainless steel bare metal stents (BMS) or paclitaxel-eluting stents (PES) were implanted in male Sprague-Dawley rats (BW 400 ± 20 g). Two and four weeks after stent implantation, the aorta were collected, fixed with 2% glutaraldehyde, and cut into two segments. One segment was used for scanning electron microscopy analysis to evaluate re-endothelialization, and the other segment was used to calculate the neointimal area. At 2 weeks after stenting, the appearance of neointimal hyperplasia was less in the PES group than in the BMS group. At 4 weeks after stenting, no significant difference in neointimal hyperplasia was observed between two groups. On the other hand, the PES group showed more thrombus formation and less re-endothelialization compared to the BMS group. This study demonstrated the ability of a novel rat model of aorta stenting via a common carotid artery to measure the efficacy and safety of commercially available drug-eluting stents.
|
|