Basic Research

HOME > R&D Achievement > Basic Research > Article
 
DATE : 21-06-10 12:07
Comparison between Basic and Inverse Dual Drug and Peptide-coated Stents in a Porcine Restenosis Model
 WRITER : stent
HIT : 772  
   B107.+Comparison+between+Basic+and+Inverse+Dual+Drug+and+Peptide-coated+Stents~.pdf (2.8M) [0] DATE : 2021-06-10 12:07:02
Eun-Jae Jang, So-Youn Lee, In-Ho Bae, Dae Sung Park, Myung Ho Jeong, Jun-Kyu Park

Comparison between Basic and Inverse Dual Drug and Peptide-coated Stents in a Porcine Restenosis Model 

Journal of Chemistry and Applied Chemical Engineering

Dual drug-eluting stents (DES) is a primary treatment method for coronary arterial diseases in current interventional cardiology practice. However, their pathological results according to the sequence of coating of drugs have not been reported yet. The peptide-dopamine dissolved in acetonitrile was coated onto the Chonnam National University Hospital (CNUH) stent using an electrospinning coating machine. For secondary coating (e.g., sirolimus coating, designated as SPS), sirolimus (SRL) and poly lactic-glycolic acid (PLGA) were mixed in tetrahydrofuran (THF), and the solution was then coated on the CNUH stent that had underwent
 the primary peptide coating using an electrospinning and spray technique. Next, the peptide-dopamine was coated on the SRL-PLGA coated stent (PSS). In this study, it was confirmed that endothelialization was prooted without being significantly affected by the coating order (SPS or PSS). The sequence of drug and peptide coating may affect the development of restenosis and PSS was effective in the prevention of restenosis compared to that of using SPS.